National Vulnerability Database

National Vulnerability Database

National Vulnerability
Database

CVE-2015-1012 Detail

Current Description

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Source:  MITRE
Description Last Modified:  03/25/2019
View Analysis Description

Impact

CVSS v3.0 Severity and Metrics:

Base Score: 7.5 HIGH
Vector: AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N (V3 legend)
Impact Score: 3.6
Exploitability Score: 3.9


Attack Vector (AV): Network
Attack Complexity (AC): Low
Privileges Required (PR): None
User Interaction (UI): None
Scope (S): Unchanged
Confidentiality (C): High
Integrity (I): None
Availability (A): None

CVSS v2.0 Severity and Metrics:

Base Score: 5.0 MEDIUM
Vector: (AV:N/AC:L/Au:N/C:P/I:N/A:N) (V2 legend)
Impact Subscore: 2.9
Exploitability Subscore: 10.0


Access Vector (AV): Network
Access Complexity (AC): Low
Authentication (AU): None
Confidentiality (C): Partial
Integrity (I): None
Availability (A): None
Additional Information:
Allows unauthorized disclosure of information

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 Mitigation Third Party Advisory US Government Resource

Technical Details

Vulnerability Type (View All)

  • Information Leak / Disclosure (CWE-200)

Known Affected Software Configurations Switch to CPE 2.2

Configuration 1 ( hide )
 cpe:2.3:o:pfizer:lifecare_pca_infusion_system_firmware:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
5.0
Running on/with
 cpe:2.3:h:pfizer:lifecare_pca_infusion_system:-:*:*:*:*:*:*:*
     Show Matching CPE(s)


Change History

1 change record found - show changes

Quick Info

CVE Dictionary Entry:
CVE-2015-1012
NVD Published Date:
03/25/2019
NVD Last Modified:
04/04/2019