CVE-2015-3956 Detail

Current Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Source:  MITRE
View Analysis Description

Severity



CVSS 3.x Severity and Metrics:

NIST CVSS score
NIST: NVD
Base Score: 9.8 CRITICAL
Vector:  CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 Mitigation Third Party Advisory US Government Resource

Weakness Enumeration

CWE-ID CWE Name Source
CWE-345 Insufficient Verification of Data Authenticity NIST   ICS-CERT  

Known Affected Software Configurations Switch to CPE 2.2

Configuration 1 ( hide )
 cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
13.4
Running on/with
 cpe:2.3:h:pifzer:plum_a\+_infusion_system:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 2 ( hide )
 cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
13.6
Running on/with
 cpe:2.3:h:pifzer:plum_a\+3_infusion_system:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 3 ( hide )
 cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
3.13
Running on/with
 cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*
     Show Matching CPE(s)


Change History

2 change records found - show changes

Quick Info

CVE Dictionary Entry:
CVE-2015-3956
NVD Published Date:
03/25/2019
NVD Last Modified:
10/09/2019