National Vulnerability Database

National Vulnerability Database

National Vulnerability
Database

CVE-2017-12726 Detail

Current Description

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.

Source:  MITRE
View Analysis Description

Impact

CVSS v3.0 Severity and Metrics:

Base Score: 7.3 HIGH
Vector: AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L (V3 legend)
Impact Score: 3.4
Exploitability Score: 3.9


Attack Vector (AV): Network
Attack Complexity (AC): Low
Privileges Required (PR): None
User Interaction (UI): None
Scope (S): Unchanged
Confidentiality (C): Low
Integrity (I): Low
Availability (A): Low

CVSS v2.0 Severity and Metrics:

Base Score: 7.5 HIGH
Vector: (AV:N/AC:L/Au:N/C:P/I:P/A:P) (V2 legend)
Impact Subscore: 6.4
Exploitability Subscore: 10.0


Access Vector (AV): Network
Access Complexity (AC): Low
Authentication (AU): None
Confidentiality (C): Partial
Integrity (I): Partial
Availability (A): Partial
Additional Information:
Allows unauthorized disclosure of information
Allows unauthorized modification
Allows disruption of service

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
http://www.securityfocus.com/bid/100665 Third Party Advisory VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A Third Party Advisory US Government Resource

Technical Details

Vulnerability Type (View All)

  • Use of Hard-coded Credentials (CWE-798)

Known Affected Software Configurations Switch to CPE 2.2

Configuration 1 ( hide )
 cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:*
     Show Matching CPE(s)
 cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.5:*:*:*:*:*:*:*
     Show Matching CPE(s)
 cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.6:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:-:*:*:*:*:*:*:*
     Show Matching CPE(s)


Change History

2 change records found - show changes

Quick Info

CVE Dictionary Entry:
CVE-2017-12726
NVD Published Date:
02/15/2018
NVD Last Modified:
03/02/2018