Current Description

Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Source:  MITRE
View Analysis Description

Impact

CVSS v3.0 Severity and Metrics:

Base Score: 5.3 MEDIUM
Vector: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N (V3 legend)
Impact Score: 3.6
Exploitability Score: 1.6


Attack Vector (AV): Adjacent
Attack Complexity (AC): High
Privileges Required (PR): None
User Interaction (UI): None
Scope (S): Unchanged
Confidentiality (C): None
Integrity (I): High
Availability (A): None

CVSS v2.0 Severity and Metrics:

Base Score: 2.9 LOW
Vector: (AV:A/AC:M/Au:N/C:N/I:P/A:N) (V2 legend)
Impact Subscore: 2.9
Exploitability Subscore: 5.5


Access Vector (AV): Local_Network
Access Complexity (AC): Medium
Authentication (AU): None
Confidentiality (C): None
Integrity (I): Partial
Availability (A): None
Additional Information:
Allows unauthorized modification

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
http://www.securityfocus.com/bid/105044 Third Party Advisory VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 Third Party Advisory US Government Resource

Technical Details

Vulnerability Type (View All)

Known Affected Software Configurations Switch to CPE 2.2

Configuration 1 ( hide )
 cpe:2.3:o:medtronicdiabetes:508_minimed_insulin_pump_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:508_minimed_insulin_pump:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 2 ( hide )
 cpe:2.3:o:medtronicdiabetes:522_paradigm_real-time_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:522_paradigm_real-time:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 3 ( hide )
 cpe:2.3:o:medtronicdiabetes:722_paradigm_real-time_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:722_paradigm_real-time:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 4 ( hide )
 cpe:2.3:o:medtronicdiabetes:523_paradigm_revel_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:523_paradigm_revel:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 5 ( hide )
 cpe:2.3:o:medtronicdiabetes:723_paradigm_revel_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:723_paradigm_revel:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 6 ( hide )
 cpe:2.3:o:medtronicdiabetes:523k_paradigm_revel_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:523k_paradigm_revel:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 7 ( hide )
 cpe:2.3:o:medtronicdiabetes:723k_paradigm_revel_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:723k_paradigm_revel:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 8 ( hide )
 cpe:2.3:o:medtronicdiabetes:551_minimed_530g_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:551_minimed_530g:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 9 ( hide )
 cpe:2.3:o:medtronicdiabetes:751_minimed_530g_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronicdiabetes:751_minimed_530g:-:*:*:*:*:*:*:*
     Show Matching CPE(s)


Change History

3 change records found - show changes

Quick Info

CVE Dictionary Entry:
CVE-2018-14781
NVD Published Date:
08/13/2018
NVD Last Modified:
10/26/2018