Description

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.

Source:  MITRE
Description Last Modified:  11/20/2018

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
http://www.securityfocus.com/bid/105843
https://ics-cert.us-cert.gov/advisories/ICSMA-18-310-01