Current Description

Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system.

Source:  MITRE
View Analysis Description

Impact

CVSS v3.0 Severity and Metrics:

Base Score: 7.8 HIGH
Vector: AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H (V3 legend)
Impact Score: 5.9
Exploitability Score: 1.8


Attack Vector (AV): Local
Attack Complexity (AC): Low
Privileges Required (PR): Low
User Interaction (UI): None
Scope (S): Unchanged
Confidentiality (C): High
Integrity (I): High
Availability (A): High

CVSS v2.0 Severity and Metrics:

Base Score: 7.2 HIGH
Vector: (AV:L/AC:L/Au:N/C:C/I:C/A:C) (V2 legend)
Impact Subscore: 10.0
Exploitability Subscore: 3.9


Access Vector (AV): Local
Access Complexity (AC): Low
Authentication (AU): None
Confidentiality (C): Complete
Integrity (I): Complete
Availability (A): Complete
Additional Information:
Allows unauthorized disclosure of information
Allows unauthorized modification
Allows disruption of service

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
http://www.securityfocus.com/bid/104088 Third Party Advisory VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01 Third Party Advisory US Government Resource
https://www.usa.philips.com/healthcare/about/customer-support/product-security Vendor Advisory

Technical Details

Vulnerability Type (View All)

  • Use of Hard-coded Credentials (CWE-798)

Known Affected Software Configurations Switch to CPE 2.2

Configuration 1 ( hide )
 cpe:2.3:o:philips:brilliance_firmware_64:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
2.6.2
Running on/with
 cpe:2.3:h:philips:brilliance_64:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 2 ( hide )
 cpe:2.3:o:philips:brilliance_ict_sp_firmware:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
3.2.4
Running on/with
 cpe:2.3:h:philips:brilliance_ict_sp:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 3 ( hide )
 cpe:2.3:o:philips:brilliance_ict_firmware:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
4.1.6
Running on/with
 cpe:2.3:h:philips:brilliance_ict:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 4 ( hide )
 cpe:2.3:o:philips:_brilliance_ct_big_bore_firmware:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
2.3.5
Running on/with
 cpe:2.3:h:philips:_brilliance_ct_big_bore:-:*:*:*:*:*:*:*
     Show Matching CPE(s)


Change History

2 change records found - show changes

Quick Info

CVE Dictionary Entry:
CVE-2018-8857
NVD Published Date:
05/04/2018
NVD Last Modified:
06/13/2018