National Vulnerability Database

National Vulnerability Database

National Vulnerability
Database

CVE-2018-8868 Detail

Current Description

Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Source:  MITRE
Description Last Modified:  07/02/2018
View Analysis Description

Impact

CVSS v3.0 Severity and Metrics:

Base Score: 6.4 MEDIUM
Vector: AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H (V3 legend)
Impact Score: 5.9
Exploitability Score: 0.5


Attack Vector (AV): Physical
Attack Complexity (AC): High
Privileges Required (PR): None
User Interaction (UI): None
Scope (S): Unchanged
Confidentiality (C): High
Integrity (I): High
Availability (A): High

CVSS v2.0 Severity and Metrics:

Base Score: 6.9 MEDIUM
Vector: (AV:L/AC:M/Au:N/C:C/I:C/A:C) (V2 legend)
Impact Subscore: 10.0
Exploitability Subscore: 3.4


Access Vector (AV): Local
Access Complexity (AC): Medium
Authentication (AU): None
Confidentiality (C): Complete
Integrity (I): Complete
Availability (A): Complete
Additional Information:
Allows unauthorized disclosure of information
Allows unauthorized modification
Allows disruption of service

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01 Third Party Advisory US Government Resource

Technical Details

Vulnerability Type (View All)

  • Improper Access Control (CWE-284)

Change History