National Vulnerability Database

National Vulnerability Database

National Vulnerability
Database

CVE-2019-6540 Detail

Current Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.

Source:  MITRE
View Analysis Description

Impact

CVSS v3.0 Severity and Metrics:

Base Score: 6.5 MEDIUM
Vector: AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N (V3.0 legend)
Impact Score: 3.6
Exploitability Score: 2.8


Attack Vector (AV): Adjacent
Attack Complexity (AC): Low
Privileges Required (PR): None
User Interaction (UI): None
Scope (S): Unchanged
Confidentiality (C): High
Integrity (I): None
Availability (A): None

CVSS v2.0 Severity and Metrics:

Base Score: 3.3 LOW
Vector: (AV:A/AC:L/Au:N/C:P/I:N/A:N) (V2 legend)
Impact Subscore: 2.9
Exploitability Subscore: 6.5


Access Vector (AV): Local_Network
Access Complexity (AC): Low
Authentication (AU): None
Confidentiality (C): Partial
Integrity (I): None
Availability (A): None
Additional Information:
Allows unauthorized disclosure of information

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
http://www.securityfocus.com/bid/107544 Third Party Advisory VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01 Third Party Advisory US Government Resource

Technical Details

Vulnerability Type (View All)

Known Affected Software Configurations Switch to CPE 2.2

Configuration 1 ( hide )
 cpe:2.3:o:medtronic:mycarelink_monitor_24950_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:mycarelink_monitor_24950:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 2 ( hide )
 cpe:2.3:o:medtronic:mycarelink_monitor_24952_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:mycarelink_monitor_24952:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 3 ( hide )
 cpe:2.3:o:medtronic:carelink_monitor_2490c_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:carelink_monitor_2490c:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 4 ( hide )
 cpe:2.3:o:medtronic:carelink_2090_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:carelink_2090:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 5 ( hide )
 cpe:2.3:o:medtronic:amplia_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:amplia_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 6 ( hide )
 cpe:2.3:o:medtronic:claria_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:claria_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 7 ( hide )
 cpe:2.3:o:medtronic:compia_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:compia_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 8 ( hide )
 cpe:2.3:o:medtronic:concerto_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:concerto_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 9 ( hide )
 cpe:2.3:o:medtronic:concerto_ii_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:concerto_ii_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 10 ( hide )
 cpe:2.3:o:medtronic:consulta_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:consulta_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 11 ( hide )
 cpe:2.3:o:medtronic:evera_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:evera_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 12 ( hide )
 cpe:2.3:o:medtronic:maximo_ii_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:maximo_ii_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 13 ( hide )
 cpe:2.3:o:medtronic:maximo_ii_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:maximo_ii_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 14 ( hide )
 cpe:2.3:o:medtronic:mirro_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:mirro_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 15 ( hide )
 cpe:2.3:o:medtronic:nayamed_nd_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:nayamed_nd_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 16 ( hide )
 cpe:2.3:o:medtronic:primo_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:primo_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 17 ( hide )
 cpe:2.3:o:medtronic:protecta_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:protecta_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 18 ( hide )
 cpe:2.3:o:medtronic:protecta_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:protecta_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 19 ( hide )
 cpe:2.3:o:medtronic:secura_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:secura_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 20 ( hide )
 cpe:2.3:o:medtronic:virtuoso_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:virtuoso_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 21 ( hide )
 cpe:2.3:o:medtronic:virtuoso_ii_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:virtuoso_ii_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 22 ( hide )
 cpe:2.3:o:medtronic:visia_af_icd_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:visia_af_icd:-:*:*:*:*:*:*:*
     Show Matching CPE(s)

Configuration 23 ( hide )
 cpe:2.3:o:medtronic:viva_crt-d_firmware:-:*:*:*:*:*:*:*
     Show Matching CPE(s)
Running on/with
 cpe:2.3:h:medtronic:viva_crt-d:-:*:*:*:*:*:*:*
     Show Matching CPE(s)


Change History

3 change records found - show changes

Quick Info

CVE Dictionary Entry:
CVE-2019-6540
NVD Published Date:
03/26/2019
NVD Last Modified:
10/09/2019