Description
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Metrics
NVD enrichment efforts reference publicly available information to associate
vector strings. CVSS information contributed by other sources is also
displayed.
By selecting these links, you will be leaving NIST webspace.
We have provided these links to other web sites because they
may have information that would be of interest to you. No
inferences should be drawn on account of other sites being
referenced, or not, from this page. There may be other web
sites that are more appropriate for your purpose. NIST does
not necessarily endorse the views expressed, or concur with
the facts presented on these sites. Further, NIST does not
endorse any commercial products that may be mentioned on
these sites. Please address comments about this page to [email protected].
Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
CVE Modified by ICS-CERT1/31/2025 12:15:16 PM
Action
Type
Old Value
New Value
Changed
Description
The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
New CVE Received from ICS-CERT1/30/2025 2:15:14 PM
Action
Type
Old Value
New Value
Added
Description
The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.