Search Parameters:
  • CPE Product Version: cpe:/h:hospira:lifecare_pca3:-
There are 6 matching records.
Displaying matches 1 through 6.
Vuln ID Summary CVSS Severity
CVE-2015-3958

Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets.

Published: July 06, 2015; 3:59:04 PM -0400
V3.x:(not available)
V2.0: 7.8 HIGH
CVE-2015-3957

Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.

Published: July 06, 2015; 3:59:03 PM -0400
V3.x:(not available)
V2.0: 4.6 MEDIUM
CVE-2015-3955

Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.

Published: July 06, 2015; 3:59:02 PM -0400
V3.x:(not available)
V2.0: 10.0 HIGH
CVE-2015-1011

Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors.

Published: July 06, 2015; 3:59:01 PM -0400
V3.x:(not available)
V2.0: 5.0 MEDIUM
CVE-2014-5406

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

Published: July 06, 2015; 3:59:00 PM -0400
V3.x:(not available)
V2.0: 9.3 HIGH
CVE-2015-3459

The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.

Published: April 29, 2015; 7:59:00 PM -0400
V3.x:(not available)
V2.0: 10.0 HIGH