You are viewing this page in an unauthorized frame window.
This is a potential security issue, you are being redirected to
https://nvd.nist.gov
An official website of the United States government
Official websites use .gov
A .gov website belongs to an official government organization in the United States.
Secure .gov websites use HTTPS
A lock () or https:// means you've safely connected to the .gov website. Share sensitive information only on official, secure websites.
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
Metrics
NVD enrichment efforts reference publicly available information to associate
vector strings. CVSS information contributed by other sources is also
displayed.
By selecting these links, you will be leaving NIST webspace.
We have provided these links to other web sites because they
may have information that would be of interest to you. No
inferences should be drawn on account of other sites being
referenced, or not, from this page. There may be other web
sites that are more appropriate for your purpose. NIST does
not necessarily endorse the views expressed, or concur with
the facts presented on these sites. Further, NIST does not
endorse any commercial products that may be mentioned on
these sites. Please address comments about this page to nvd@nist.gov.
El módulo de comunicación en el Hospira LifeCare PCA Infusion System en versiones anteriores a 7.0 no requiere autenticación para sesiones TELNET root, lo que permite a atacantes remotos modificar la configuración de la bomba a través de comandos no especificados.
Removed
Translation
La bomba de infusión de Hospira Lifecare PCA funcionado con 'SW ver 412' no requiere autenticación para las sesiones Telnet, lo que permite a atacantes remotos ganar privilegios root a través de un puerto TCP 23.
Modified Analysis by NIST7/13/2015 3:36:17 PM
Action
Type
Old Value
New Value
Changed
CPE Configuration
Configuration 1
AND
OR
*cpe:2.3:o:hospira:lifecare_pcainfusion_pump_firmware:412:*:*:*:*:*:*:*
OR
cpe:2.3:h:hospira:lifecare_pcainfusion_pump:*:*:*:*:*:*:*:*
Configuration 1
AND
OR
*cpe:2.3:o:hospira:lifecare_pcainfusion_firmware:5.0:*:*:*:*:*:*:* (and previous)
OR
*cpe:2.3:h:hospira:lifecare_pca5:-:*:*:*:*:*:*:*
*cpe:2.3:h:hospira:lifecare_pca3:-:*:*:*:*:*:*:*
Changed
Reference Type
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm US Govt Resource
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm Advisory, US Govt Resource
Changed
Reference Type
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 US Govt Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 Advisory, US Govt Resource
CVE Modified by MITRE7/08/2015 9:59:06 PM
Action
Type
Old Value
New Value
Changed
Description
Hospira Lifecare PCA infusion pump running "SW ver 412" does not require authentication for Telnet sessions, which allows remote attackers to gain root privileges via TCP port 23.
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
Configuration 1
AND
OR
*cpe:2.3:o:hospira:lifecare_pcainfusion_pump_firmware:412:*:*:*:*:*:*:*
OR
cpe:2.3:h:hospira:lifecare_pcainfusion_pump:*:*:*:*:*:*:*:*
Added
CVSS V2
(AV:N/AC:L/Au:N/C:C/I:C/A:C)
Added
CWE
CWE-264
Initial CVE Analysis4/30/2015 2:06:05 PM
Action
Type
Old Value
New Value
Quick Info
CVE Dictionary Entry: CVE-2015-3459 NVD
Published Date: 04/29/2015 NVD
Last Modified: 01/03/2017
Source: MITRE