Current Description

In the most recent firmware for Blipcare, the device provides an open Wireless network called "Blip" for communicating with the device. The user connects to this open Wireless network and uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is in vicinity of Wireless signal generated by the Blipcare device to easily sniff the credentials. Also, an attacker can connect to the open wireless network "Blip" exposed by the device and modify the HTTP response presented to the user by the device to execute other attacks such as convincing the user to download and execute a malicious binary that would infect a user's computer or mobile device with malware.

Source:  MITRE
View Analysis Description

Impact

CVSS v3.0 Severity and Metrics:

Base Score: 7.1 HIGH
Vector: AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:N (V3 legend)
Impact Score: 4.2
Exploitability Score: 2.8


Attack Vector (AV): Adjacent
Attack Complexity (AC): Low
Privileges Required (PR): None
User Interaction (UI): None
Scope (S): Unchanged
Confidentiality (C): High
Integrity (I): Low
Availability (A): None

CVSS v2.0 Severity and Metrics:

Base Score: 4.8 MEDIUM
Vector: (AV:A/AC:L/Au:N/C:P/I:P/A:N) (V2 legend)
Impact Subscore: 4.9
Exploitability Subscore: 6.5


Access Vector (AV): Local_Network
Access Complexity (AC): Low
Authentication (AU): None
Confidentiality (C): Partial
Integrity (I): Partial
Availability (A): None
Additional Information:
Allows unauthorized disclosure of information
Allows unauthorized modification

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
http://packetstormsecurity.com/files/153225/Blipcare-Clear-Text-Communication-Memory-Corruption.html Third Party Advisory VDB Entry
https://github.com/ethanhunnt/IoT_vulnerabilities/blob/master/Blipcare_sec_issues.pdf Exploit Third Party Advisory
https://seclists.org/bugtraq/2019/Jun/8 Mailing List Third Party Advisory

Technical Details

Vulnerability Type (View All)

  • 7PK - Security Features (CWE-254)

Known Affected Software Configurations Switch to CPE 2.2

Configuration 1 ( hide )
 cpe:2.3:o:blipcare:wi-fi_blood_pressure_monitor_firmware:*:*:*:*:*:*:*:*
     Show Matching CPE(s)
Up to (including)
bp700_10.1
Running on/with
 cpe:2.3:h:blipcare:wi-fi_blood_pressure_monitor:-:*:*:*:*:*:*:*
     Show Matching CPE(s)


Change History

1 change record found - show changes

Quick Info

CVE Dictionary Entry:
CVE-2017-11579
NVD Published Date:
07/02/2019
NVD Last Modified:
07/15/2019