National Vulnerability Database

National Vulnerability Database

National Vulnerability
Database

CVE-2017-9657 Detail

Current Description

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Source:  MITRE
Description Last Modified:  04/30/2018
View Analysis Description

Impact

CVSS v3.0 Severity and Metrics:

Base Score: 6.5 MEDIUM
Vector: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H (V3 legend)
Impact Score: 3.6
Exploitability Score: 2.8


Attack Vector (AV): Adjacent
Attack Complexity (AC): Low
Privileges Required (PR): None
User Interaction (UI): None
Scope (S): Unchanged
Confidentiality (C): None
Integrity (I): None
Availability (A): High

CVSS v2.0 Severity and Metrics:

Base Score: 3.3 LOW
Vector: (AV:A/AC:L/Au:N/C:N/I:N/A:P) (V2 legend)
Impact Subscore: 2.9
Exploitability Subscore: 6.5


Access Vector (AV): Local_Network
Access Complexity (AC): Low
Authentication (AU): None
Confidentiality (C): None
Integrity (I): None
Availability (A): Partial
Additional Information:
Allows disruption of service

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
http://www.securityfocus.com/bid/100813 Third Party Advisory VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 Third Party Advisory US Government Resource
https://www.usa.philips.com/healthcare/about/customer-support/product-security Vendor Advisory