Vuln ID Summary CVSS Severity
CVE-2019-10964

In Medtronic MinMed 508 and Medtronic Minimed Paradigm Insulin Pumps, Versions, MiniMed 508 pump – All versions, MiniMed Paradigm 511 pump – All versions, MiniMed Paradigm 512/712 pumps – All versions, MiniMed Paradigm 712E pump–All versions, MiniMed Paradigm 515/715 pumps–All versions, MiniMed Paradigm 522/722 pumps – All versions,MiniMed Paradigm 522K/722K pumps – All versions, MiniMed Paradigm 523/723 pumps – Software versions 2.4A or lower, MiniMed Paradigm 523K/723K pumps – Software, versions 2.4A or lower, MiniMed Paradigm Veo 554/754 pumps – Software versions 2.6A or lower, MiniMed Paradigm Veo 554CM and 754CM models only – Software versions 2.7A or lower, the affected insulin pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Published: June 28, 2019; 5:15:11 PM -0400
V3.0: 8.8 HIGH
V2.0: 5.8 MEDIUM
CVE-2019-6540

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.

Published: March 26, 2019; 2:29:01 PM -0400
V3.1: 6.5 MEDIUM
V2.0: 3.3 LOW
CVE-2019-6538

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Published: March 25, 2019; 6:29:00 PM -0400
V3.1: 6.5 MEDIUM
V2.0: 3.3 LOW
CVE-2018-18984

Medtronic CareLink 2090 Programmer CareLink 9790 Programmer 29901 Encore Programmer, all versions, The affected products do not encrypt or do not sufficiently encrypt the following sensitive information while at rest PII and PHI.

Published: December 14, 2018; 10:29:00 AM -0500
V3.1: 4.6 MEDIUM
V2.0: 2.1 LOW
CVE-2018-10626

A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product's update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.

Published: August 10, 2018; 2:29:00 PM -0400
V3.0: 4.4 MEDIUM
V2.0: 3.8 LOW
CVE-2018-10596

Medtronic 2090 CareLink Programmer all versions The affected product uses a virtual private network connection to securely download updates. The product does not verify it is still connected to this virtual private network before downloading updates. An attacker with local network access to the programmer could influence these communications.

Published: July 02, 2018; 9:29:00 PM -0400
V3.0: 8.0 HIGH
V2.0: 5.2 MEDIUM
CVE-2018-5448

All versions of the Medtronic 2090 Carelink Programmer are affected by a directory traversal vulnerability where the product's software deployment network could allow an attacker to read files on the system.

Published: May 04, 2018; 2:29:00 PM -0400
V3.0: 5.7 MEDIUM
V2.0: 2.7 LOW
CVE-2018-5446

All versions of the Medtronic 2090 Carelink Programmer are affected by a per-product username and password that is stored in a recoverable format which could allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network.

Published: May 04, 2018; 2:29:00 PM -0400
V3.0: 5.3 MEDIUM
V2.0: 2.1 LOW